DocumentCode
846959
Title
Trace Impurity Analyzer
Author
Schneider, W.J. ; Edwards, D., Jr.
Volume
26
Issue
3
fYear
1979
fDate
6/1/1979 12:00:00 AM
Firstpage
4091
Lastpage
4092
Abstract
The desirability for long-term reliability of large scale helium refrigerator systems used on superconducting accelerator magnets has necessitated detection of impurities to levels of a few ppm. An analyzer that measures trace impurity levels of condensable contaminants in concentrations of less than a ppm in 15 atm of He is described. The instrument makes use of the desorption temperature at an indicated pressure of the various impurities to determine the type of contaminant. The pressure rise at that temperature yields a measure of the contaminant level of the impurity. A LN2 cryogenic charcoal trap is also employed to measure air impurities (nitrogen and oxygen) to obtain the full range of contaminant possibilities. The results of this detector which will be in use on the research and development helium refrigerator of the ISABELLE First-Cell is described.
Keywords
Accelerator magnets; Helium; Impurities; Instruments; Large-scale systems; Magnetic analysis; Pollution measurement; Refrigeration; Superconducting magnets; Temperature measurement;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1979.4330709
Filename
4330709
Link To Document