• DocumentCode
    848109
  • Title

    Gamma-Ray Response of a 38-mm Bismuth Germanate Scintillator

  • Author

    Evans, Albert E., Jr.

  • Author_Institution
    University of California, Los Alamos Scientific Laboratory Los Alamos, New Mexico 87545
  • Volume
    27
  • Issue
    1
  • fYear
    1980
  • Firstpage
    172
  • Lastpage
    175
  • Abstract
    The response of a 38-mm-diam by 38-mm-long bismuth germanate (BGO) scintillator to gamma radiation from 0.12 to 8.28 MeV has been measured. The full-energypeak efficiency of the detector, which varies from 80% at 0.12 MeV to 10.8% at 8.29 MeV, is 5.6 times as great at 2.75 MeV as the full-energy efficiency of a similar NaI (Tl) detector. Although the energy resolution of the detector is inferior to that which may be obtained with NaI (Tl), its high peak-to-Compton ratio and high full-energy-to-escape-peak ratio is advantageous. This detector shows promise of application in many fields.
  • Keywords
    Bismuth; Face detection; Gamma ray detection; Gamma ray detectors; Gamma rays; NIST; Neutrons; Protons; Pulse measurements; Radiation detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1980.4330822
  • Filename
    4330822