DocumentCode
848109
Title
Gamma-Ray Response of a 38-mm Bismuth Germanate Scintillator
Author
Evans, Albert E., Jr.
Author_Institution
University of California, Los Alamos Scientific Laboratory Los Alamos, New Mexico 87545
Volume
27
Issue
1
fYear
1980
Firstpage
172
Lastpage
175
Abstract
The response of a 38-mm-diam by 38-mm-long bismuth germanate (BGO) scintillator to gamma radiation from 0.12 to 8.28 MeV has been measured. The full-energypeak efficiency of the detector, which varies from 80% at 0.12 MeV to 10.8% at 8.29 MeV, is 5.6 times as great at 2.75 MeV as the full-energy efficiency of a similar NaI (Tl) detector. Although the energy resolution of the detector is inferior to that which may be obtained with NaI (Tl), its high peak-to-Compton ratio and high full-energy-to-escape-peak ratio is advantageous. This detector shows promise of application in many fields.
Keywords
Bismuth; Face detection; Gamma ray detection; Gamma ray detectors; Gamma rays; NIST; Neutrons; Protons; Pulse measurements; Radiation detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1980.4330822
Filename
4330822
Link To Document