Reduction of Polarization in Mercuric Iodide Nuclear Radiation Detectors
Author
Holzer, A. ; Schieber, M.
Author_Institution
School of Applied Science and Technology, Hebrew University of Jerusalem Jerusalem, Israel
Volume
27
Issue
1
fYear
1980
Firstpage
266
Lastpage
271
Abstract
The majority of HgI2 detectors which have an energy resolution less than 5% for 241Am do not have severe plarization effects at all. Polarization effects are reported for detectors having energy resolution greater than 5% for 241Am in an attempt to discover if polarization is mainly due to either a nonoptimal fabrication or to severe trapping in the bulk crystalline material. An improved fabrication procedure using a rapid etch and evaporated palladium contacts is reported. This procedure has produced polarization free detectors which were refabricated from detectors which previously showed polarization.
Keywords
Crystalline materials; Crystals; Degradation; Energy resolution; Fabrication; Leak detection; Leakage current; Polarization; Radiation detectors; Voltage;