• DocumentCode
    848571
  • Title

    Computerized Longitudinal Tomography (CLT) with a Bilateral Collimator

  • Author

    Nalcioglu, O. ; Morton, M.E. ; Milne, N.

  • Author_Institution
    Department of Radiological Sciences University of California-Irvine Irvine, CA 92717
  • Volume
    27
  • Issue
    1
  • fYear
    1980
  • Firstpage
    430
  • Lastpage
    434
  • Abstract
    A nuclear Medicine imaging system which yields Longitudinal tomographic slices through an object is presented. The system consists of a dual view parallel hole (Bilateral) collimator, a gamma camera and a minicomputer. The bilateral collimator enables one to acquire two projections of the object simultaneously. After one position, the collimator is rotated with respect to the camera face to obtain another pair of projections of the object at the new angle. After a sufficient number of projections are taken, longitudinal slices through the object at any depth are reconstructed using the minicomputer interfaced to the camera. Phantom and patient data are presented to demonstrate the capabilities of the system.
  • Keywords
    Cameras; Collimators; Computed tomography; Image reconstruction; Microcomputers; Nuclear medicine; Optical imaging; Radioactive materials; Three dimensional displays; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1980.4330865
  • Filename
    4330865