DocumentCode :
849571
Title :
Non-Linear Techniques in Multi-Spectral X-Ray Imaging
Author :
Heinzerling, J. ; Schlindwein, M.
Author_Institution :
Philips GmbH Forschungslaboratorium Hamburg, Postbox 54 08 40 2000 Hamburg 54 F.R.G.
Volume :
27
Issue :
2
fYear :
1980
fDate :
4/1/1980 12:00:00 AM
Firstpage :
961
Lastpage :
968
Abstract :
The sensitivity of medical radiographs for the contrast agent iodine can be improved by multispectral imaging techniques. Three different shadow projections of the object to be imaged are taken by using three distinct primary X-ray spectra. The corresponding material-selective projections of iodine, bone and tissue masses are generated by combining appropriately the shadow projections. A nonlinear reconstruction algorithm is presented thereby improving the reconstruction accuracy compared with known linear techniques. The algorithm is tested both in computer simulation and in experiment.
Keywords :
Biological materials; Bones; Computer simulation; Electromagnetic wave absorption; Filters; Gratings; Image reconstruction; Reconstruction algorithms; Testing; X-ray imaging;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1980.4330961
Filename :
4330961
Link To Document :