DocumentCode :
851558
Title :
Semiconductor Detectors for Medical Tomography with High-Energy Heavy Ions
Author :
Llacer, Jorge ; Haller, Eugen E. ; Hansen, William L. ; Walton, John T. ; Batho, Eleanor K.
Author_Institution :
Lawrence Berkeley Laboratory University of California Berkeley, California 94720
Volume :
28
Issue :
1
fYear :
1981
Firstpage :
161
Lastpage :
166
Abstract :
High-energy heavy ion beams are in use at the Lawrence Berkeley Laboratory for cancer therapy. In order to take full advantage of the very favorable depth-dose characteristics of those beams, it is necessary to determine the stopping characteristics of the ions in the complex media of a human with greater accuracy than obtainable with X-ray CAT scanning. Initial measurements with an array of Si dE/dx position sensitive detectors and a windowless thin planar Ge detector used in a side entry mode show the potential for fabricating an instrument for high accuracy on-line CAT scanning using the same ions to be used for therapy. It is estimated that one tomography can be obtained with a dose of 0.72 Rad-gm.
Keywords :
Cancer; Humans; Instruments; Ion beams; Laboratories; Medical treatment; Position measurement; Position sensitive particle detectors; Sensor arrays; X-ray tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1981.4331159
Filename :
4331159
Link To Document :
بازگشت