Title :
Pulse Height Response of Cd-Te to Heavy Ions
Author :
Becchetti, F.D. ; Ristinen, R.A. ; Peterson, R.J. ; Hamill, J.J. ; Thorn, C.E. ; Entine, G.
Author_Institution :
University of Colorado, Boulder, CO 80309
fDate :
4/1/1981 12:00:00 AM
Abstract :
Pulse height spectra for alpha particles, fission fragments, and oxygen ions have been obtained for a thin-window surface-barrier Cd-Te detector. Although the detector can exhibit a large pulse-height defect and poorer resolution than Si detectors, it is still usable in many applications. In addition, the high stopping power of Cd-Te makes it suitable for simultaneous detection of energetic light and heavy ions, as well as ¿ rays, with a simple compact room-temperature detector.
Keywords :
Alpha particles; Conductivity; Detectors; Energy resolution; Laboratories; Particle scattering; Protons; Radiation monitoring; Temperature; Voltage;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1981.4331463