DocumentCode :
854923
Title :
Flexible transcutaneous transformer for artificial heart system
Author :
Matsuki, H. ; Shiiki, M. ; Murakami, K. ; Nadehara, K. ; Yamamoto, T.
Author_Institution :
Dept. of Electr. Eng., Tohoku Univ., Sendai, Japan
Volume :
26
Issue :
5
fYear :
1990
fDate :
9/1/1990 12:00:00 AM
Firstpage :
1548
Lastpage :
1550
Abstract :
A flexible transcutaneous transformer utilizing amorphous magnetic fiber for an artificial heart system is reported. The transformer can transmit electrical energy of 20 W with an efficiency of over 90%; the primary current density does not greatly depend on the relative displacement of the coils. The temperature rise of the transformer was about 6°C in a phantom test
Keywords :
artificial organs; cardiology; transformers; 20 W; 90 percent; amorphous magnetic fiber; artificial heart system; efficiency; flexible transcutaneous transformer; phantom test; primary current density; temperature rise; Amorphous magnetic materials; Amorphous materials; Artificial heart; Batteries; Coils; Copper; Magnetic flux; Power transformer insulation; Skin; Temperature;