DocumentCode :
855776
Title :
Guarded vector voltmeter for AC ratio standard calibration
Author :
Callegaro, Luca ; D´Elia, Vincenzo
Author_Institution :
Electr. Metrol. Dept., Ist. Elettrotecnico Nazionale Galileo Ferraris, Torino, Italy
Volume :
51
Issue :
4
fYear :
2002
fDate :
8/1/2002 12:00:00 AM
Firstpage :
632
Lastpage :
635
Abstract :
A new setup for the calibration of the ratio of inductive voltage dividers against a ratio standard is under development at IEN. The setup is based on the direct measurement of voltage vector differences between the device under test and the standard, with a purposely built vector voltmeter described here. The uniqueness of the voltmeter is the extremely high common-mode-rejection-ratio, greater than 180 dB at 1 kHz, obtained with a bootstrapping-guarding technique. The voltmeter is battery-operated, has variable gain and a dual-channel synchronous demodulating stage with high harmonic rejection, and is operated with a personal computer through an optical-fiber interface.
Keywords :
calibration; measurement standards; voltage dividers; voltmeters; 1 kHz; AC ratio standard calibration; bootstrapping-guarding technique; common-mode-rejection-ratio; dual-channel synchronous demodulation; guarded vector voltmeter; harmonic rejection; inductive voltage divider; optical fiber interface; personal computer; Calibration; Gain; Measurement standards; Microcomputers; Optical computing; Standards development; Testing; Ultraviolet sources; Voltage measurement; Voltmeters;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2002.802258
Filename :
1044675
Link To Document :
بازگشت