  • Title

    Development of semi-portable DC-SQUID magnetometer

  • Author

    Kandori, A. ; Ueda, M. ; Ogata, H. ; Kado, H.

  • Author_Institution
    Superconducting Sensor Lab, Chiba, Japan
  • Volume
    5
  • Issue
    2
  • fYear
    1995
  • fDate
    6/1/1995 12:00:00 AM
  • Firstpage
    2474
  • Lastpage
    2477
  • Abstract
    A semi-portable DC-SQUID magnetometer with a liquid helium continuous flow cryostat has been developed. The cryostat is 340 mm long and 50 mm in outer diameter. The pick-up coil is a co-axial, first order, differential type, 20 mm in diameter and 30 mm in baseline. The distance between the pick-up coil and the outer surface of the cryostat is 5 mm. It is easy to set this cryostat on the chest wall or on a human head. The cool down time is about 20 minutes, which is much shorter than that for a conventional storage Dewar. The white noise level, measured in a magnetically shielded room, is 8.7 fT//spl radic/Hz at 30 Hz. Noise at 1 Hz is about 30 fT//spl radic/Hz. Noise from temperature fluctuations (<9 mK) and drift (4.22 K /spl sim/ 4.55 K) is negligible. The consumption rate of liquid helium is 10 l/h. Here, the authors show the measurement of a magnetocardiogram in a magnetically shielded room.<>
  • Keywords
    SQUID magnetometers; biomedical equipment; cryostats; magnetocardiography; 1 Hz; 20 min; 20 to 340 mm; 5 mm; chest wall; conventional storage Dewar; cool down time; human head; liquid helium continuous flow cryostat; magnetically shielded room; magnetocardiogram measurement; medical instrumentation; pick-up coil; semiportable DC-SQUID magnetometer; temperature fluctuations; white noise level; Coils; Fluid flow; Humans; Magnetic heads; Magnetic noise; Magnetic shielding; Magnetometers; Noise measurement; White noise;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/77.403092
  • Filename