Title
Performance improvement of Si(Li) Peltier cooled detectors
Author
Sokolov, A. ; Gostilo, V. ; Loupilov, A. ; Zalinkevich, V.
Author_Institution
Baltic Sci. Instrum., Riga, Latvia
Volume
49
Issue
5
fYear
2002
fDate
10/1/2002 12:00:00 AM
Firstpage
2427
Lastpage
2430
Abstract
The spectrometric performance of Si(Li) Peltier cooled detectors was improved due to better electronics, a small-sized vacuum chamber, and an effective five-channel Peltier cooler, which provided a temperature difference of 132°C at a sensitivity to heat loading that is less than 15 mWt/°C. This allowed us to fabricate circular Si(Li) Peltier cooled detectors with a thickness of 4-4.5 mm and a sensitive surface from 12 to 100 mm2. Their energy resolution at an energy of 5.9 keV was from 158 to 253 eV depending on the sensitive surface area, while on 59.6 keV it was from 410 to 460 eV. Rectangular detectors with a thickness 4-4.5 mm and sensitive surface area 60 mm2 had an energy resolution of 255 and 450 eV for energies of 5.9 and 59.6 keV, respectively.
Keywords
Peltier effect; position sensitive particle detectors; silicon radiation detectors; 4 to 4.5 mm; 5.9 keV; 59.6 keV; Peltier cooled detector; Si(Li) detector; Si:Li; energy resolution; heat loading; sensitive surface area; thickness; vacuum chamber; Energy resolution; Fabrication; Fluorescence; Leakage current; Radiation detectors; Silicon radiation detectors; Temperature; Thermoelectric devices; X-ray detection; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2002.803861
Filename
1046760
Link To Document