Title :
Recent Developments in Thick Mercuric Iodide Spectrometers
Author :
Hull, K. ; Beyerle, A. ; Lopez, B. ; Markakis, J. ; Ortale, C. ; Schnepple, W. ; van den Berg, L.
Author_Institution :
EG&G, Santa Barbara Operations 130 Robin Hill Road Goleta, California 93116
Abstract :
Thick (~l cm) mercuric iodide gamma-ray detectors have been produced which show spectroscopic qualities at moderate detector biases (~5 kV) comparable to those of thin spectrometers. Efficiency measurements indicate that the entire volume of the detectors is active. Spectra resolutions of less than 10% have been obtained for gamma-ray energies above 1 MeV. Short charge collection times have produced the best results. Measurement of crystal charge transport properties is discussed. A small amount of bias conditioning is necessary for best performance. Operating parameters of the detectors have been investigated.
Keywords :
Charge measurement; Counting circuits; Crystals; Current measurement; Gamma ray detection; Gamma ray detectors; Signal resolution; Signal to noise ratio; Spectroscopy; Volume measurement;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1983.4332299