• DocumentCode
    863503
  • Title

    Esophageal pulse oximetry utilizing reflectance photoplethysmography

  • Author

    Kyriacou, Panayiotis A. ; Powell, Sarah ; Langford, Richard M. ; Jones, Deric P.

  • Author_Institution
    Dept. of Eng., Univ. of London, UK
  • Volume
    49
  • Issue
    11
  • fYear
    2002
  • Firstpage
    1360
  • Lastpage
    1368
  • Abstract
    Peripheral perfusion is often poor and barely pulsatile in patients undergoing prolonged major surgery. Hence, the arterial blood oxygen saturation (SpO2) readings from commercial finger pulse oximeters can become unreliable or cease when they are most needed. To overcome this limitation, the esophagus has been investigated as an alternative measurement site, as perfusion may be preferentially preserved centrally. A reflectance esophageal pulse oximeter probe, and a processing system implemented in Lab VIEW were developed. The system was evaluated in clinical measurements on 49 cardiothoracic surgery patients. The SpO2 values from the esophagus were in good agreement with arterial blood oxygen saturation (SaO2) values obtained from blood gas analysis and CO-oximetry. The means (±SD) of the differences between the esophageal SpO2 and SaO2 results from blood gas analysis and CO-oximetry were 0.02±0.88% and -0.73±0.72%, respectively. In five (10.2%) of the patients, the finger pulse oximeter failed for at least 10 min while the esophageal SpO2 readings remained reliable. The results confirm that the esophagus may be used as an alternative monitoring site for pulse oximetry even in patients with compromised peripheral perfusion.
  • Keywords
    biological organs; haemorheology; oximetry; patient monitoring; plethysmography; probes; reflectivity; surgery; Lab VIEW-based processing system; O/sub 2/; alternative monitoring site; arterial blood oxygen saturation; cardiothoracic surgery patients; commercial finger pulse oximeters; compromised peripheral perfusion; esophageal pulse oximetry; finger pulse oximeter; peripheral perfusion; prolonged major surgery; reflectance esophageal pulse oximeter probe; reflectance photoplethysmography; Blood; Cardiology; Condition monitoring; Esophagus; Fingers; Hospitals; Physics; Reflectivity; Standards publication; Surgery; Adult; Aged; Aged, 80 and over; Equipment Design; Equipment Failure Analysis; Esophagus; Female; Humans; Male; Middle Aged; Oximetry; Oxygen; Photoplethysmography; Reproducibility of Results; Sensitivity and Specificity;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2002.804584
  • Filename
    1046944