Title :
A Single-Stage Photomultiplier/Amplifier Combination for Use in Intense Magnetic Fields
Author :
Rousseau, M.D. ; Brown, R.M. ; Jeffreys, P. ; Edwards, M. ; Sproston, M.
Author_Institution :
Rutherford Appleton Laboratory, Chilton, Didcot, Oxon, England
Abstract :
The use of a single-stage photomultiplier followed by a low-noise amplifier has made it possible to obtain cosmic ray muon pulse height spectra from F2 type lead glass with a resolution closely comparable to that achieved with a conventional photomultiplier. Tests in axial magnetic fields show satisfactory behaviour over the range 0.05 Tesla to at least 1 Tesla. A further similar device designed for transverse field use has operated up to 0.8 Tesla at angles approaching 700 to the device longitudinal axis.
Keywords :
Detectors; Glass; Laboratories; Low-noise amplifiers; Magnetic fields; Mesons; Photomultipliers; Prototypes; Pulse amplifiers; Testing;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :