Abstract :
This paper describes a pocket-calculator-sized, neutron-sensitive, REM-responding personnel dosimeter that uses three tissue-equivalent cylindrical proportional counters as neutron-sensitive detectors. These are conventionally called Linear Energy Transfer (LET) counters. Miniaturized hybrid circuits are used for the linear pulse handling electronics, followed by a 256-channel ADC. A CMOS microprocessor is used to calculate REM exposure from the basic rads-tissue data supplied by the LET counters and also to provide timing and display functions. The instrument is used to continuously accumulate time in hours since reset, total counts accumulated, rads-tissue, and REM. The user can display any one of these items or a channel number (an aid in calibration) at any time. Such data are provided with a precision of ±3% for a total exposure of 1 mREM over eight hours.
Keywords :
Counting circuits; Detectors; Displays; Energy exchange; Instruments; Microprocessors; Neutrons; Personnel; Pulse circuits; Timing;