Abstract :
The digital imaging of carbohydrate consumption in myocardial tissue has been performed by using a deposit tracer of glucose metabolism (3H-deoxyglucose) and a specially designed Multiwire Proportional Chamber (MWPC) that acts as a position sensitive radioactivity detector. The resolving power (¿1.5 mm FWHM), sensitivity (¿10-1Bq/cm2), efficiency (¿10%) and uniformity (±4%) of the system are shown and MWPC digital autoradiographs of canine heart in different pathophysiological conditions are presented.
Keywords :
Absorption; Biochemistry; Cathodes; Delay; Detectors; Digital images; Heart; Myocardium; Spatial resolution; Sugar;