DocumentCode :
863986
Title :
A Comparison of the Noise Characteristics of Projection Reconstruction and Two-Dimensional Fourier Transformations in NMR Imaging
Author :
Ortendahl, D.A. ; Crooks, L.E. ; Kaufman, L.
Author_Institution :
University of California, San Francisco Radiologic Imaging Laboratory 400 Grandview Drive South San Francisco, CA 94080
Volume :
30
Issue :
1
fYear :
1983
Firstpage :
692
Lastpage :
696
Abstract :
Human images of excellent quality have been obtained with Nuclear Magnetic Resonance (NMR), using two different techniques: projection reconstruction (PR) and two-dimensional Fourier transformations (2DFT). It has been shown in CT that the reconstruction algorithms will multiply the noise in the projections and change the noise spectrum. A comparison of the noise of the two techniques is presented under the constraint of equal imaging conditions using both computer simulations and data obtained with the UCSF human imager. The computer simulations show that the two techniques produce approximately equal signal to noise values when S/N ~ 15. Below that value small differences are observed. For both simulations and actual data, the 2DFT shows a uniform spatial frequency response while the PR method shows a roll-off at low spatial frequencies. The 2DFT method is shown to be more resistant to artifact formation.
Keywords :
Computational modeling; Computed tomography; Computer simulation; Frequency response; Humans; Image reconstruction; Magnetic noise; Magnetic resonance imaging; Nuclear magnetic resonance; Reconstruction algorithms;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1983.4332357
Filename :
4332357
Link To Document :
بازگشت