• DocumentCode
    864017
  • Title

    Engineering Developments on the UBC-TRIUMF Modified PETT VI Positron Emission Tomograph

  • Author

    Evans, B. ; Harrop, R. ; Heywood, D. ; Mackintosh, J. ; Moore, R.W. ; Pate, B.D. ; Rogers, J.G. ; Ruth, T.J. ; Sayre, C. ; Sprenger, H. ; Van Oers, N. ; Guang, Yao Xiao

  • Author_Institution
    TRIUMF, University of British Columbia, Vancouver, B.C., V6T 2A3, Canada
  • Volume
    30
  • Issue
    1
  • fYear
    1983
  • Firstpage
    707
  • Lastpage
    710
  • Abstract
    A tomograph with the PETT VI geometry has been built with improvements generally applicable to such devices. In addition to improved temperature control, the gantry features commercial CsF detectors with the newer Amperex photomultiplier tubes (PMT\´s). Much of the coincidence support circuitry is of an original design, utilizing new "fast" TTL family devices. A local DEC 11/23 microprocessor provides for routine diagnostic and reliability checking, gantry control, and acquisition of single detector counting rates. Image reconstruction and display is performed by a medium size VAX 11/780 computer, operating in a time-sharing environment. Some preliminary performance characteristics of the machine have been measured. The reconstructed resolution in-slice, as well as the reconstructed slice thickness, has been measured as a function of radius for both "straight" and "cross" slices.
  • Keywords
    Circuits; Computer displays; Computer vision; Detectors; Geometry; Image reconstruction; Photomultipliers; Radioactive decay; Temperature control; Time sharing computer systems;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1983.4332361
  • Filename
    4332361