• DocumentCode
    866147
  • Title

    Calibration of an Analyzing Magnet Using a Time-of-Flight Technique

  • Author

    Lennard, W.N. ; Phillips, D. ; Hill, R.

  • Author_Institution
    Atomic Energy of Canada Limited Research Company Solid State Science Branch Chalk River Nuclear Laboratories Chalk River, Ontario, Canada KOJ 1JO
  • Volume
    30
  • Issue
    2
  • fYear
    1983
  • fDate
    4/1/1983 12:00:00 AM
  • Firstpage
    1532
  • Lastpage
    1534
  • Abstract
    We have calibrated the analyzing magnet (Danfysik 90° sector, double-focussing, 2 m radius) serving the CRNL 2 MV High Voltage Mass Separator. The method depends on a measurement of the time-of-flight, and thus of the velocity, of magnetically analyzed ions of known mass and charge state. Results have been obtained for ions in the mass range 1-208 amu for fields 0.4-12.8 kG.
  • Keywords
    Calibration; Charge measurement; Current measurement; Energy measurement; Magnetic analysis; Magnetic field measurement; Magnetic separation; Particle separators; Saturation magnetization; Voltage;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1983.4332578
  • Filename
    4332578