Title :
The Fine Grid Detector : A Parallel Electrode Postion Sensitive Detector
Author_Institution :
European Molecular Biology Laboratory, Notkestrasse 85, 2000 Hanburg 52, West Germany
Abstract :
A parallel electrode type of position sensitive detector with a fast delay-line read-out system has been developed. As a preliminary result, a spatial resolution of 0,5mm FWHM has been obtained. The counting rate capability is as high as 105 sec-1 mm-2.
Keywords :
Anodes; Electrodes; Electromagnetic wave absorption; Electrons; Laboratories; Position sensitive particle detectors; Spatial resolution; Wires; X-ray detection; X-ray detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1984.4333262