DocumentCode
873119
Title
Energy Resolution Enhancement of Mercuric Iodide Detectors
Author
Finger, M. ; Prince, T.A. ; Padgett, L. ; Prickett, B. ; Schnepple, W.
Author_Institution
California Institute of Technology, Pasadena, CA 91125
Volume
31
Issue
1
fYear
1984
Firstpage
348
Lastpage
352
Abstract
A pulse processing technique has been developed which improves the gamma-ray energy resolution of mercuric iodide detectors. The technique employs a fast (100 ns) and a slow (6.4 ¿s) pulse height analysis to correct for signal variations due to variations in charge trapping. The capabilities of the technique for energy resolution enhancement are discussed as well as the utility of the technique for examining the trapping characteristics of individual detectors. We have achieved an energy resolution of 2.6% FWHM at 662 keV with an acceptance efficiency of 100% from a mercuric iodide detector which gives 8.3% FWHM using standard techniques.
Keywords
Charge carrier processes; Electrodes; Electron traps; Energy resolution; Signal analysis; Signal resolution; Temperature; Time measurement; X-ray detection; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1984.4333274
Filename
4333274
Link To Document