DocumentCode :
873825
Title :
An on-Line Radio-Nuclide Liquid Effluent Monitor with a Reduced Contamination Rate
Author :
Malm, H.L. ; Walker, R.A.P. ; Mallory, P.G.
Author_Institution :
Aptec Engineering Ltd. 4251 Steeles Avenue West Downsview, Ontario M3N 1V7
Volume :
31
Issue :
1
fYear :
1984
Firstpage :
668
Lastpage :
671
Abstract :
A problem with on-line monitoring of radionuclides in liquid effluent is the build-up of contamination and background on all wetted surfaces. A combination of beta monitoring, a dual phosphor scintillator, and sample geometry was used to reduce the rate of increase of background compared to other types of monitors. Performance in a nuclear station with an effluent (before dilution) of 1.7MBq/m3 (45uCi/m3) Cs 137 equivalent showed a contamination rate of 800 times less than a monitor without the above features.
Keywords :
Containers; Detectors; Effluents; Geometry; Monitoring; Particle scattering; Phosphors; Pollution measurement; Power engineering and energy; Surface contamination;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1984.4333342
Filename :
4333342
Link To Document :
بازگشت