Title :
Handheld-impedance-measurement system with seven-decade capability and potentiostatic function
Author :
Angelini, Emma ; Carullo, Alessio ; Corbellini, Simone ; Ferraris, Franco ; Gallone, Vincenzo ; Grassini, Sabrina ; Parvis, Marco ; Vallan, Alberto
Author_Institution :
Dipt. dei Mater.i ed Ingegneria Chimica, Politecnico di Torino, Italy
fDate :
4/1/2006 12:00:00 AM
Abstract :
This paper describes design and test of a new impedance-measurement system for nonlinear devices that exhibits a seven-decade range and works down to a frequency of 0.01 Hz. The system is specifically designed for electrochemical measurements, but the proposed architecture can be employed in many other fields where flexible signal generation and analysis are required. The system employs an unconventional signal generator based on two pulsewidth modulation (PWM) oscillators and an autocalibration system that allows uncertainties of less than 3% to be obtained over a range of 1 kΩ to 100 GΩ. A synchronous demodulation processing allows the noise superimposed to the low-amplitude input signals to be made negligible.
Keywords :
electric impedance measurement; oscillators; portable instruments; signal generators; 0.01 Hz; autocalibration system; corrosion testing; electrochemical devices; electrochemical measurements; flexible signal generation; handheld impedance measurement system; intelligent systems; nonlinear devices; potentiostatic function; pulsewidth modulation oscillators; signal analysis; signal processing; synchronous demodulation processing; Demodulation; Frequency measurement; Impedance measurement; Oscillators; Pulse width modulation; Signal analysis; Signal design; Signal generators; Space vector pulse width modulation; System testing; Corrosion testing; electrochemical devices; impedance measurement; intelligent systems; signal processing;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
DOI :
10.1109/TIM.2006.870103