DocumentCode :
880236
Title :
Machinable Rogowski coil, design, and calibration
Author :
Ramboz, John D.
Author_Institution :
RAMTech Eng. Inc., Clermont, FL, USA
Volume :
45
Issue :
2
fYear :
1996
fDate :
4/1/1996 12:00:00 AM
Firstpage :
511
Lastpage :
515
Abstract :
Rogowski coils are special types of mutual inductors often used to measure high AC and transient currents. Traditional designs are reviewed. The significant sources of error associated with typical coil designs are examined. A “machinable Rogowski coil” is introduced and discussed. The reasoning behind critical design choices is discussed. The improved performance of the machinable coil is compared to that of other designs. Test and calibration methods are discussed
Keywords :
calibration; coils; electric current measurement; electric sensing devices; inductors; AC current measurement; calibration methods; coil design; critical design choices; current sensors; edge test; linearity; machinable Rogowski coil; mutual inductors; off-centre rotational test; pancake test; sources of error; transient current measurement; Calibration; Coils; Conductors; Current measurement; Inductance; Inductors; Instruments; Noise measurement; Shape; Voltage measurement;