Theoretical Comparison of the Iodine Quantification Accuracy of Two Spectral CT Technologies
Author :
Bornefalk, Hans ; Persson, Mats
Author_Institution :
Dept. of Phys., AlbaNova Univ. Center, Stockholm, Sweden
Volume :
33
Issue :
2
fYear :
2014
fDate :
Feb. 2014
Firstpage :
556
Lastpage :
565
Abstract :
We compare the theoretical limits of iodine quantification for the photon counting multibin and dual energy technologies. Dual energy systems by necessity have to make prior assumptions in order to quantify iodine. We explicitly allow the multibin system to make the same assumptions and also allow them to be wrong. We isolate the effect of technology from imperfections and implementation issues by assuming both technologies to be ideal, i.e., without scattered radiation, unity detection efficiency and perfect energy response functions, and by applying the Cramér-Rao lower bound methodology to assess the quantification accuracy. When priors are wrong the maximum likelihood estimates will be biased and the mean square error of the quantification error is a more appropriate figure of merit. The evaluation assumes identical X-ray spectra for both methodologies and for that reason a sensitivity analysis is performed with regard to the assumed X-ray spectrum. We show that when iodine is quantified over regions of interest larger than 6 cm2, multibin systems benefit by independent estimation of three basis functions. For smaller regions of interest multibin systems can increase quantification accuracy by making the same prior assumptions as dual energy systems.
Keywords :
X-ray spectra; computerised tomography; image reconstruction; iodine; maximum likelihood estimation; mean square error methods; medical image processing; photon counting; sensitivity analysis; Cramer-Rao lower bound methodology; dual energy systems; dual energy technologies; figure-of-merit; identical X-ray spectra; image reconstruction; iodine quantification accuracy; maximum likelihood estimates; mean square error; photon counting multibin; quantification error; regions-of-interest; sensitivity analysis; spectral computerised tomography technologies; theoretical comparison; Biological tissues; Computed tomography; Detectors; Maximum likelihood estimation; Mean square error methods; Photonics; Standards; Dual energy computed tomography (CT); iodine quantification; photon counting multibin CT; spectral CT;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2013.2290198
Filename :
6658951
Link To Document :
بازگشت