• DocumentCode
    886980
  • Title

    Adaptive control of multiplexed closed-circuit anesthesia

  • Author

    Jee, Gyu-In ; Roy, Rob J.

  • Author_Institution
    Dept. of Biomed. Eng., Rensselaer Polytech. Inst., Troy, NY, USA
  • Volume
    39
  • Issue
    10
  • fYear
    1992
  • Firstpage
    1071
  • Lastpage
    1080
  • Abstract
    The design of an adaptive closed-circuit anesthesia controller based on a multiplexed mass spectrometer system is described. The controller deals with measurement deterioration caused by measurement delay and rise time through a long catheter as well as long sampling times due to the multiplexed measurements. Measurement data are extrapolated between sampling periods to increase the estimation convergence rate. A multiple-step-ahead predictive control algorithm is used to calculate intermediate control inputs between sampling intervals. Simulations are used to validate the designed controller.
  • Keywords
    adaptive control; chemical variables control; control system synthesis; mass spectroscopic chemical analysis; patient monitoring; predictive control; adaptive control; anesthesia controller; design; estimation convergence rate; long catheter; long sampling times; measurement delay; measurement deterioration; multiple-step-ahead predictive control algorithm; multiplexed closed-circuit anesthesia; multiplexed mass spectrometer system; rise time; sampling periods; Adaptive control; Anesthesia; Catheters; Control systems; Delay effects; Mass spectroscopy; Programmable control; Sampling methods; Time measurement; Weight control; Algorithms; Anesthesia, Closed-Circuit; Equipment Design; Humans; Mass Spectrometry; Monitoring, Intraoperative; Pulmonary Gas Exchange; Signal Processing, Computer-Assisted;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.161339
  • Filename
    161339