DocumentCode :
887011
Title :
Clinical application of an active electrode using an operational amplifier
Author :
Nishimura, Suzushi ; Tomita, Yutaka ; Horiuchi, Toshio
Author_Institution :
Inst. of Biomed. Eng., Keio Univ., Kanagawa, Japan
Volume :
39
Issue :
10
fYear :
1992
Firstpage :
1096
Lastpage :
1099
Abstract :
An active electrode (d10 mm, t6 mm) that functions as an impedance transformer (with an input impedance >10 G Omega and an output impedance <1 Omega ) by means of which one can derive surface EMG without any skin preparation and paste is presented. This electrode was compared with a conventional one, and it was ascertained that the electrode could replace the conventional one and was preferable because it required less preparation time and was less affected by environmental noise.
Keywords :
bioelectric potentials; biomedical electronics; biomedical equipment; microelectrodes; muscle; operational amplifiers; 10 mm; 6 mm; active electrode; clinical application; environmental noise; impedance transformer; operational amplifier; preparation time; surface EMG; Biomedical electrodes; Biomedical measurements; Electromyography; Muscles; Operational amplifiers; Pain; Skin; Surface impedance; Surface resistance; Surface topography; Amplifiers; Electric Impedance; Electrodes; Electrophysiology; Equipment Design; Humans;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.161342
Filename :
161342
Link To Document :
بازگشت