Title :
Clinical application of an active electrode using an operational amplifier
Author :
Nishimura, Suzushi ; Tomita, Yutaka ; Horiuchi, Toshio
Author_Institution :
Inst. of Biomed. Eng., Keio Univ., Kanagawa, Japan
Abstract :
An active electrode (d10 mm, t6 mm) that functions as an impedance transformer (with an input impedance >10 G Omega and an output impedance <1 Omega ) by means of which one can derive surface EMG without any skin preparation and paste is presented. This electrode was compared with a conventional one, and it was ascertained that the electrode could replace the conventional one and was preferable because it required less preparation time and was less affected by environmental noise.
Keywords :
bioelectric potentials; biomedical electronics; biomedical equipment; microelectrodes; muscle; operational amplifiers; 10 mm; 6 mm; active electrode; clinical application; environmental noise; impedance transformer; operational amplifier; preparation time; surface EMG; Biomedical electrodes; Biomedical measurements; Electromyography; Muscles; Operational amplifiers; Pain; Skin; Surface impedance; Surface resistance; Surface topography; Amplifiers; Electric Impedance; Electrodes; Electrophysiology; Equipment Design; Humans;
Journal_Title :
Biomedical Engineering, IEEE Transactions on