DocumentCode
890712
Title
Measuring the temperature dependence of resistivity of high purity copper using a solenoid coil (SRPM method)
Author
Nakane, Hiroshi ; Watanabe, Tuneo ; Nagata, Chohjyu ; Fujiwara, Satoshi ; Yoshizawa, Shuji
Author_Institution
Sci. Univ. of Tokyo, Japan
Volume
41
Issue
1
fYear
1992
fDate
2/1/1992 12:00:00 AM
Firstpage
107
Lastpage
110
Abstract
The resistivity of high purity copper was measured by a method which estimates it by using the difference in the impedance between a circular multilayer solenoid coil with a circular rod-shaped copper sample and a similar coil without a copper sample (SRPM method). The residual resistivity ratio (RRR) of high purity copper measured at 100 Hz by the SRPM method has correlated well with the values measured by the DC four-probe method. It was confirmed that an accurate measurement of the resistivity to 10-12 Ωm is possible by the SRPM method. Frequency dependence was confirmed to exist in high purity copper with very low resistivity. As the measuring frequency is raised, the decrease in skin depth seems to affect the resistivity
Keywords
copper; electrical conductivity measurement; solenoids; 100 Hz; Cu; circular multilayer solenoid coil; electrical resistivity measurement; residual resistivity ratio; rod; skin depth; temperature dependence of resistivity; Coils; Conductivity; Copper; Frequency dependence; Frequency measurement; Impedance measurement; Nonhomogeneous media; Solenoids; Temperature dependence; Temperature measurement;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.126642
Filename
126642
Link To Document