Title :
Charge injection limits of activated iridium oxide electrodes with 0.2 ms pulses in bicarbonate buffered saline (neurological stimulation application)
Author :
Beebe, Xenia ; Rose, T.L.
Author_Institution :
EIC Labs. Inc., Norwood, MA, USA
fDate :
6/1/1988 12:00:00 AM
Abstract :
The maximum charge injection without electrode solution decomposition for activated iridium wire electrodes in bicarbonate buffered saline was 2.1 and 1.0 mC/cm2 geometric for anodic-first and cathodic-first, respectively, 0.2-ms balanced-charge biphasic current pulses. Electrodes biased at +0.8 V vs. SCE (saturated calomel electrode) accepted charge up to 3.5 mC/cm2 geometric with monophasic cathodal pulses.
Keywords :
biomedical equipment; electrodes; iridium compounds; balanced-charge biphasic current pulses; bicarbonate buffered saline; monophasic cathodal pulses; Area measurement; Counting circuits; Electrodes; Hydrogen; In vivo; Microelectrodes; Proteins; Surface cleaning; Testing; Wire; Bicarbonates; Buffers; Electric Conductivity; Electrodes; Iridium;
Journal_Title :
Biomedical Engineering, IEEE Transactions on