Abstract :
The chief possible causes of error in the calibration of microwave noise sources are examined and a radiometer designed to minimize these errors is described. This includes a self-contained hot-load standard, facilities for measuring the hot-guide losses, a synchronously switched broad-band circulator, a switchable, high-stability attenuator, provision for measuring this attenuation, and a carbon-filament lamp reference noise source.
Keywords :
Attenuation measurement; Attenuators; Australia; Calibration; Laboratories; Lamps; Loss measurement; Measurement standards; Microwave radiometry; Noise measurement;