    In this paper, a quick and efficient approach is proposed to estimate in vivo RF-induced heating from small medical implants under MRI procedure. An efficient methodology is developed to correlate the incident electric field to the induced temperature rise for stents of arbitrary shapes. A 10-cm-long titanium rod and a 6-cm-long intestine stent are used as examples to validate the approach under in vitro testing conditions. In vivo temperature rise estimations are then performed on an anatomically corrected adult male model. Numerical and experimental studies demonstrate the efficiency and accuracy of the proposed method.
  • Keywords
    biomedical MRI; medical image processing; numerical analysis; MRI procedure; estimate MRI RF field; incident electric field; induced temperature; small medical implants; vivo heating; Heating; Implants; In vivo; Phantoms; Radio frequency; Temperature; Temperature measurement; In vivo heating estimation; MRI safety; RF-induced heating; medical implants; scaling approach;