DocumentCode
8935
Title
An Efficient Approach to Estimate MRI RF Field Induced In Vivo Heating for Small Medical Implants
Author
Dawei Li ; Jianfeng Zheng ; Yan Liu ; Changwang Pan ; Kainz, Wolfgang ; Fan Yang ; Wen Wu ; Ji Chen
Author_Institution
Dept. of Electr. & Comput. Eng., Univ. of Houston, Houston, TX, USA
Volume
57
Issue
4
fYear
2015
fDate
Aug. 2015
Firstpage
643
Lastpage
650
Abstract
In this paper, a quick and efficient approach is proposed to estimate in vivo RF-induced heating from small medical implants under MRI procedure. An efficient methodology is developed to correlate the incident electric field to the induced temperature rise for stents of arbitrary shapes. A 10-cm-long titanium rod and a 6-cm-long intestine stent are used as examples to validate the approach under in vitro testing conditions. In vivo temperature rise estimations are then performed on an anatomically corrected adult male model. Numerical and experimental studies demonstrate the efficiency and accuracy of the proposed method.
Keywords
biomedical MRI; medical image processing; numerical analysis; MRI procedure; estimate MRI RF field; incident electric field; induced temperature; small medical implants; vivo heating; Heating; Implants; In vivo; Phantoms; Radio frequency; Temperature; Temperature measurement; In vivo heating estimation; MRI safety; RF-induced heating; medical implants; scaling approach;
fLanguage
English
Journal_Title
Electromagnetic Compatibility, IEEE Transactions on
Publisher
ieee
ISSN
0018-9375
Type
jour
DOI
10.1109/TEMC.2015.2435519