• DocumentCode
    896539
  • Title

    Image Reconstruction and Noise Evaluation in Photon Time-of-Flight Assisted Positron Emission Tomography

  • Author

    Tomitani, Takehiro

  • Author_Institution
    Physics Division National Institute of Radiological Sciences 9-1, Anagawa-4-Chome, Chiba-shi, 260, Japan
  • Volume
    28
  • Issue
    6
  • fYear
    1981
  • Firstpage
    4581
  • Lastpage
    4589
  • Abstract
    In positron CT, the path difference of annhilation pair gamma rays can be measured by time-of-flight (TOF) difference of pair gamma rays. This TOF information gives us rough position information along a projection line and will reduce noise propagation in the reconstruction process. A reconstruction algorithm for TOF-positron computed tomography (PCT) based on the back-projection with 1-dimensional weight and 2-dimensional filtering is presented. Also a formula to evaluate the variance of the reconstructed image and the optimal back-projection function are presented. The advantage of TOF-PCT over conventional PCT was investigated in view of noise figure. An example of such noise figure evaluations for CsF and liquid Xenon scintillators is given.
  • Keywords
    Computed tomography; Filtering; Gamma rays; Image reconstruction; Noise figure; Noise reduction; Positron emission tomography; Reconstruction algorithms; Single photon emission computed tomography; Xenon;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1981.4335769
  • Filename
    4335769