DocumentCode :
897073
Title :
Count Rate Performance of a Microchannel Plate Photomultiplier
Author :
Nieschmidt, E.B. ; Lawrence, R.S. ; Gentillon, C.D. ; Vegors, S.H., Jr.
Author_Institution :
EG&G Idaho, Inc. P. O. Box 1625 Idaho Falls, ID 83415
Volume :
29
Issue :
1
fYear :
1982
Firstpage :
196
Lastpage :
199
Abstract :
The count rate dependent gain change of a microchannel plate photomultiplier has been measured. Gain changes of 60% have been observed for count rates in excess of 2 × 105 s-1. The microchannel plate photomultiplier was used as a multiplying element of a scintillation detector system using NaI(Tl) as the scintillator. Data are compared to a simplified model of this form of photomultiplier.
Keywords :
Cathodes; Conductivity; Gain measurement; Microchannel; Photomultipliers; Preamplifiers; Pulse amplifiers; Pulse measurements; Pulsed power supplies; Scintillation counters;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1982.4335826