• DocumentCode
    897226
  • Title

    Instrumentation Reliability Study for Future Waste Repository Sites

  • Author

    Rogue, F.

  • Author_Institution
    Lawrence Livermore National Laboratory University of California Livermore, California 94550
  • Volume
    29
  • Issue
    1
  • fYear
    1982
  • Firstpage
    264
  • Lastpage
    266
  • Abstract
    In evaluating the reliability of instruments for waste repository applications, we found geotechnical instruments to be the least reliable. Geotechnical instruments measure the engineering aspects of geology, geophysics, and hydrology. Systems were designed with off-the-shelf instrumentation which was designed for short-term, ambient temperature measurements. Reliability problems developed when these off-the-shelf instruments were used in long-term high-temperature DOE experiments. The Stripa, Climax, and Avery Island DOE experiments had identical instrumentation failures with this off-the-shelf design approach. The DOE experimenters, working with instrument manufacturers, have been successful in obtaining more reliable instruments. However, issues must still be addressed if high reliability is to be achieved at future high-level waste repository sites.
  • Keywords
    Geology; Geophysical measurements; Geophysics; Hydrologic measurements; Hydrology; Instruments; Manufacturing; Reliability engineering; Temperature measurement; US Department of Energy;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1982.4335841
  • Filename
    4335841