Pocket Neutron REM Meter
Author
Quam, W. ; Duca, T. Del ; Plake, W. ; Graves, G. ; DeVore, T.
Author_Institution
EG&G, Santa Barbara Operations 130 Robin Hill Road Goleta, California 93116
Volume
29
Issue
1
fYear
1982
Firstpage
637
Lastpage
640
Abstract
This paper describes a pocket-calculator-sized, neutron-sensitive, REM-responding personnel dosimeter that uses two tissue-equivalent cylindrical proportional counters as neutron-sensitive detectors. These are conventionally called Linear Energy Transfer (LET) counters. Miniaturized hybrid circuits are used for the linear pulse handling electronics, followed by a 256-channel ADC. A CMOS microprocessor is used to calculate REM exposure from the basic rads-tissue data supplied by the LET counters and also to provide timing and display functions. The instrument is used to continuously accumulate time in hours since reset, total counts accumulated, rads-tissue, and REM. At any time the user can display any one of these items or a channel number (an aid in calibration). The instrument provides such data with a precision of +3% for a total exposure of 1 mREM over 8 hours.
Keywords
Counting circuits; Detectors; Displays; Energy exchange; Instruments; Microprocessors; Neutrons; Personnel; Pulse circuits; Timing;