DocumentCode :
898464
Title :
Beta-Gated Gamma Coincidence Counting with a Phoswich Detector
Author :
Kaye, J.H. ; Warner, R.A.
Author_Institution :
Pacific Northwest Laboratory P. 0. Box 999 Richland, Washington 99352
Volume :
29
Issue :
1
fYear :
1982
Firstpage :
812
Lastpage :
815
Abstract :
A special type of phoswich detector system has been evaluated for measurement of radionuclides which decay with emission of time coincident beta and gamma radiation. Background reductions of more than two orders of magnitude have been obtained for the energy region from 500-950 keV. Both NE 102 plastic scintillators and anthracene were evaluated. Advantages and disadvantages of the method are discussed.
Keywords :
Coaxial cables; Face detection; Gamma ray detection; Gamma ray detectors; Photomultipliers; Plastics; Pulse circuits; Pulse measurements; Pulse shaping methods; Radiation detectors;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1982.4335965
Filename :
4335965
Link To Document :
بازگشت