• DocumentCode
    900455
  • Title

    Dielectric measurements with an open-ended coaxial probe

  • Author

    Marsland, T.P. ; Evans, S.

  • Author_Institution
    Cambridge University, Engineering Department, Cambridge, UK
  • Volume
    134
  • Issue
    4
  • fYear
    1987
  • fDate
    8/1/1987 12:00:00 AM
  • Firstpage
    341
  • Lastpage
    349
  • Abstract
    The paper is concerned with the use of an open-ended coaxial-line probe for making dielectric measurements on lossy materials. The input reflection coefficient of the probe is measured with a network analyser, and the permittivity is determined using an approximate circuit model for the probe admittance. Applying the cross-ratio invariance property of the bilinear transform to two such circuit models, we derive a simplified procedure that permits simultaneous calibration of the probe and error correction of the network analyser. Estimates of the resulting uncertainties are given, and the technique is illustrated with experimental measurements on known dielectric liquids between 50 MHz and 2.6 GHz.
  • Keywords
    calibration; dielectric measurement; microwave measurement; network analysers; permittivity measurement; probes; 50 MHz to 2.6 GHz; bilinear transform; calibration; cross-ratio invariance property; dielectric liquids; dielectric measurements; error correction; input reflection coefficient; lossy materials; network analyser; open-ended coaxial-line probe; permittivity; probe admittance;
  • fLanguage
    English
  • Journal_Title
    Microwaves, Antennas and Propagation, IEE Proceedings H
  • Publisher
    iet
  • ISSN
    0950-107X
  • Type

    jour

  • DOI
    10.1049/ip-h-2.1987.0068
  • Filename
    4643028