DocumentCode
900455
Title
Dielectric measurements with an open-ended coaxial probe
Author
Marsland, T.P. ; Evans, S.
Author_Institution
Cambridge University, Engineering Department, Cambridge, UK
Volume
134
Issue
4
fYear
1987
fDate
8/1/1987 12:00:00 AM
Firstpage
341
Lastpage
349
Abstract
The paper is concerned with the use of an open-ended coaxial-line probe for making dielectric measurements on lossy materials. The input reflection coefficient of the probe is measured with a network analyser, and the permittivity is determined using an approximate circuit model for the probe admittance. Applying the cross-ratio invariance property of the bilinear transform to two such circuit models, we derive a simplified procedure that permits simultaneous calibration of the probe and error correction of the network analyser. Estimates of the resulting uncertainties are given, and the technique is illustrated with experimental measurements on known dielectric liquids between 50 MHz and 2.6 GHz.
Keywords
calibration; dielectric measurement; microwave measurement; network analysers; permittivity measurement; probes; 50 MHz to 2.6 GHz; bilinear transform; calibration; cross-ratio invariance property; dielectric liquids; dielectric measurements; error correction; input reflection coefficient; lossy materials; network analyser; open-ended coaxial-line probe; permittivity; probe admittance;
fLanguage
English
Journal_Title
Microwaves, Antennas and Propagation, IEE Proceedings H
Publisher
iet
ISSN
0950-107X
Type
jour
DOI
10.1049/ip-h-2.1987.0068
Filename