• DocumentCode
    901649
  • Title

    Effects of low charge injection densities on corrosion responses of pulsed 316LVM stainless steel electrodes [functional neuromuscular stimulation]

  • Author

    Riedy, Lisa W. ; Walter, James S.

  • Author_Institution
    Rehabilitation Res. & Dev. Center, Hines VA Hospital, IL, USA
  • Volume
    43
  • Issue
    6
  • fYear
    1996
  • fDate
    6/1/1996 12:00:00 AM
  • Firstpage
    660
  • Lastpage
    663
  • Abstract
    The safe charge injection density for pulsing of 316LVM electrodes has been reported to be 40 μC/cm 2. However, only 20 μC/cm 2 is available for nonfaradic charge transfer and double layer charge injection. Therefore, the authors evaluated long term pulsing at 20 μC/cm 2 with capacitor coupling.
  • Keywords
    bioelectric phenomena; charge-coupled devices; corrosion; muscle; neurophysiology; prosthetics; stainless steel; capacitor coupling; corrosion responses; double layer charge injection; functional neuromuscular stimulation; implanted biomedical electrodes; long term pulsing; low charge injection densities effects; nonfaradic charge transfer; pulsed 316LVM stainless steel electrodes; Capacitors; Charge transfer; Corrosion; Electrodes; Hospitals; In vitro; Proteins; Research and development; Steel; Wire; Albumins; Animals; Cattle; Corrosion; Electric Conductivity; Electric Stimulation; Electric Stimulation Therapy; Electrodes; Evaluation Studies as Topic; Materials Testing; Microscopy, Electron, Scanning; Stainless Steel; Surface Properties;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/10.495287
  • Filename
    495287