DocumentCode :
906035
Title :
Phase-error-free quadrature sampling technique in the ultrasonic B-scan imaging system and its application to the synthetic focusing system
Author :
Chang, Seong Ho ; Park, Song Bai ; Cho, Gyu Hyoung
Author_Institution :
Dept. of Electr. Eng., Korea Adv. Inst. of Sci. & Technol., Seoul, South Korea
Volume :
40
Issue :
3
fYear :
1993
fDate :
5/1/1993 12:00:00 AM
Firstpage :
216
Lastpage :
223
Abstract :
The quadrature sampling technique as a means of detecting the envelope of RF waveform in the baseband is well known. If this technique is applied to a focused ultrasound imaging system using an array transducer, whether it is a synthetic or nonsynthetic focusing system, unwanted phase terms appear in the expressions of the inphase and quadrature components of the baseband signal when an appropriate delay time is introduced to each channel signal for the purpose of focusing. The expressions of the inphase and quadrature components from the point of focusing are derived and analyzed in detail, and a scheme to eliminate the unwanted phase terms is proposed. The resulting phase-error-free quadrature sampling technique is applied to the synthetic focusing system; a system block diagram together with the simulation and experimental results are presented.<>
Keywords :
acoustic arrays; acoustic imaging; acoustic signal processing; biomedical ultrasonics; focusing; PEFQST; RF waveform envelope detection; array transducer; delay time; focused ultrasound imaging system; phase-error-free quadrature sampling technique; simulation; synthetic focusing system; ultrasonic B-scan imaging system; Baseband; Delay effects; Envelope detectors; Focusing; Phased arrays; Radio frequency; Sampling methods; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/58.216834
Filename :
216834
Link To Document :
بازگشت