907093
Title
Analytical Study of Performance of High Resolution Positron Emission Computed Tomographs for Animal Study
Author
Nohara, N. ; Tanaka, E. ; Tomitani, T. ; Yamamoto, M. ; Murayama, H.
Author_Institution
Division of Physics National Institute of Radiological Sciences 9-1, Anagawa-4-chome, Chiba-shi 260, Japan
Volume
32
Issue
1
fYear
1985
Firstpage
818
Lastpage
821
Abstract
Performance of positron emission computed tomographs having spatial resolution as high as 3 mm FWHM was analytically evaluated. Increases in widths of line spread functions due to positron range and angular deviation were estimated. The slice thickness should be thin in accordance with the resolution in transverse axial plane. Singles rate, true coincidence rate, and scattered coincidence rate with high resolution tomographs of small diameter detector rings were estimated for three types of thin slice collimators. Feasibility of high resolution positron tomographs with small detector rings was discussed with applications to animal study.
Keywords
Animals; Collimators; Detectors; High performance computing; Image reconstruction; Performance analysis; Physics computing; Positrons; Radioactive decay; Spatial resolution;