• DocumentCode
    907133
  • Title

    A One-Dimensional Multigated Time-of-Flight Acquisition System

  • Author

    Hutchins, G.D. ; Votaw, J.R. ; Funk, K.M. ; Rosenthal, M.S. ; Koeppe, R.A. ; Nickles, R.J.

  • Author_Institution
    1530 Medical Sciencs Center 1300 University Avenue Madison, WI 53706
  • Volume
    32
  • Issue
    1
  • fYear
    1985
  • Firstpage
    835
  • Lastpage
    838
  • Abstract
    A multigated time-of-flight probe system has been developed for the simultaneous measurement of spatial and temporal modulations in the activity distribution of positron emitting radiotracers within the human body. This system consists of a pair of 4×4 cm BaF2 scintillators coupled to Amperex XP2020 Quartz photomultiplier tubes. Constant fraction discrimination provides the time pick-off of each anode signal required to start and stop a Time-to-Amplitude Converter. A slow overlap coincidence channel is used to define the dynamic range of the system and gate the TAC signal through to an ADC/LSI 11/02 microcomputer pair for data storage. A pulsewatch (LED-phototransistor pair) and the appropriate gating logic have been developed to incorporate gating capabilities into the system. A multigated TOF system can provide noninvasive alternatives to measuring blood tracer concentrations in positron emission tomography.
  • Keywords
    Anodes; Dynamic range; Humans; Large scale integration; Logic; Memory; Microcomputers; Photomultipliers; Positrons; Probes;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4336951
  • Filename
    4336951