DocumentCode
907133
Title
A One-Dimensional Multigated Time-of-Flight Acquisition System
Author
Hutchins, G.D. ; Votaw, J.R. ; Funk, K.M. ; Rosenthal, M.S. ; Koeppe, R.A. ; Nickles, R.J.
Author_Institution
1530 Medical Sciencs Center 1300 University Avenue Madison, WI 53706
Volume
32
Issue
1
fYear
1985
Firstpage
835
Lastpage
838
Abstract
A multigated time-of-flight probe system has been developed for the simultaneous measurement of spatial and temporal modulations in the activity distribution of positron emitting radiotracers within the human body. This system consists of a pair of 4Ã4 cm BaF2 scintillators coupled to Amperex XP2020 Quartz photomultiplier tubes. Constant fraction discrimination provides the time pick-off of each anode signal required to start and stop a Time-to-Amplitude Converter. A slow overlap coincidence channel is used to define the dynamic range of the system and gate the TAC signal through to an ADC/LSI 11/02 microcomputer pair for data storage. A pulsewatch (LED-phototransistor pair) and the appropriate gating logic have been developed to incorporate gating capabilities into the system. A multigated TOF system can provide noninvasive alternatives to measuring blood tracer concentrations in positron emission tomography.
Keywords
Anodes; Dynamic range; Humans; Large scale integration; Logic; Memory; Microcomputers; Photomultipliers; Positrons; Probes;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1985.4336951
Filename
4336951
Link To Document