Title : 
Collimator Design Using Ray-Tracing Techniques
         
        
            Author : 
Beck, Robert N. ; Redtung, Leo D.
         
        
            Author_Institution : 
The Franklin McLean Memorial Research Institute the University of Chicago Chicago, Illinois 60637
         
        
        
        
        
        
        
            Abstract : 
A new computer program, RAYTRC, has been written to analyze and design collimators for high-energy gamma-ray cameras used in nuclear medicine. The program finds collimator sensitivity, and geometric and penetration point source response functions. The results of this program are presented for a number of commercial collimators and for a new experimental design used to image with 511 keV gamma-rays. Collimators designed for such high-energy gamma-rays often produce bothersome hole pattern in the images due to thick septa. In order to analyze this problem we introduce a ´generalized collimator transfer function´ which characterizes not only the averaged collimator response, but also the effect of hole pattern on the images.
         
        
            Keywords : 
Cameras; Design for experiments; Image analysis; Lattices; Nuclear medicine; Optical collimators; Packaging; Pattern analysis; Ray tracing; Transfer functions;
         
        
        
            Journal_Title : 
Nuclear Science, IEEE Transactions on
         
        
        
        
        
            DOI : 
10.1109/TNS.1985.4336956