• DocumentCode
    907190
  • Title

    Collimator Design Using Ray-Tracing Techniques

  • Author

    Beck, Robert N. ; Redtung, Leo D.

  • Author_Institution
    The Franklin McLean Memorial Research Institute the University of Chicago Chicago, Illinois 60637
  • Volume
    32
  • Issue
    1
  • fYear
    1985
  • Firstpage
    865
  • Lastpage
    869
  • Abstract
    A new computer program, RAYTRC, has been written to analyze and design collimators for high-energy gamma-ray cameras used in nuclear medicine. The program finds collimator sensitivity, and geometric and penetration point source response functions. The results of this program are presented for a number of commercial collimators and for a new experimental design used to image with 511 keV gamma-rays. Collimators designed for such high-energy gamma-rays often produce bothersome hole pattern in the images due to thick septa. In order to analyze this problem we introduce a ´generalized collimator transfer function´ which characterizes not only the averaged collimator response, but also the effect of hole pattern on the images.
  • Keywords
    Cameras; Design for experiments; Image analysis; Lattices; Nuclear medicine; Optical collimators; Packaging; Pattern analysis; Ray tracing; Transfer functions;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1985.4336956
  • Filename
    4336956