• DocumentCode
    908847
  • Title

    Avalanche Diode Low Energy X-Ray and Nuclear Particle Detector

  • Author

    Squillante, M.R. ; Farrell, R. ; Lund, J.C. ; Sinclair, F. ; Entine, G. ; Keller, K.R.

  • Author_Institution
    Radiation Monitoring Devices, Inc. 44 Hunt Street Watertown, MA 02172
  • Volume
    33
  • Issue
    1
  • fYear
    1986
  • Firstpage
    336
  • Lastpage
    339
  • Abstract
    Avalanche diodes which are sensitive enough to detect low energy x-rays and beta particles have been fabricated and tested. They operate at room temperature and their structure is based on large area avalanche photodiodes. X-ray response has been measured over the energy range of 4 keV to 26 keV. At 5.9 keV a FWHM of 10% was obtained. In addition, multi-element sensors have been fabricated. Beta particles were detected from a variety of isotopic sources including C1-36, Sr-90, C-14 and H-3. The tritium beta particles, with 5 keV average energy, were detected with a 20% counting efficiency from a calibrated source. Tritium beta particles from biological samples were also measured. The high sensitivity to low energy x-rays and beta particles makes this detector useful for many industrial, military and commercial applications, and in particular for biomedical research.
  • Keywords
    Avalanche photodiodes; Beta rays; Biomedical measurements; Diodes; Energy measurement; Radiation detectors; Temperature sensors; Testing; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1986.4337113
  • Filename
    4337113