    Low Energy X-Ray Spectra Measured with a Mercuric Iodide Energy Dispersive Spectrometer in a Scanning Electron Microscope

  • Author

    Iwanczyk, J.S. ; Dabrowski, A.J. ; Huth, G.C. ; Bradley, J.G. ; Conley, J.M. ; Albee, A.L.

  • Author_Institution
    University of Southern California Institute of Physics 4676 Admiralty Way, Suite 932 Marina del Rey, CA 90292
  • Volume
    33
  • Issue
    1
  • fYear
    1986
  • Firstpage
    355
  • Lastpage
    358
  • Abstract
    A mercuric iodide energy dispersive x-ray spectrometer, with Peltier cooling provided for the detector and input field effect transistor, has been developed and tested in a scanning electron microscope. X-ray spectra were obtained with the 15 keV electron beam. An energy resolution of 225 eV (FWHM) for Mn-K¿ at 5.9 keV and 195 eV (FWHM) for Mg-K line at 1.25 keV has been measured. Overall system noise level was 175 eV (FWHM). The detector system characterization with a carbon target demonstrated good energy sensitivity at low energies and lack of significant spectral artifacts at higher energies.
  • Keywords
    Cooling; Dispersion; Electron beams; Energy measurement; FETs; Scanning electron microscopy; Spectroscopy; Testing; X-ray detection; X-ray detectors;