Low Energy X-Ray Spectra Measured with a Mercuric Iodide Energy Dispersive Spectrometer in a Scanning Electron Microscope
Author
Iwanczyk, J.S. ; Dabrowski, A.J. ; Huth, G.C. ; Bradley, J.G. ; Conley, J.M. ; Albee, A.L.
Author_Institution
University of Southern California Institute of Physics 4676 Admiralty Way, Suite 932 Marina del Rey, CA 90292
Volume
33
Issue
1
fYear
1986
Firstpage
355
Lastpage
358
Abstract
A mercuric iodide energy dispersive x-ray spectrometer, with Peltier cooling provided for the detector and input field effect transistor, has been developed and tested in a scanning electron microscope. X-ray spectra were obtained with the 15 keV electron beam. An energy resolution of 225 eV (FWHM) for Mn-K¿ at 5.9 keV and 195 eV (FWHM) for Mg-K line at 1.25 keV has been measured. Overall system noise level was 175 eV (FWHM). The detector system characterization with a carbon target demonstrated good energy sensitivity at low energies and lack of significant spectral artifacts at higher energies.
Keywords
Cooling; Dispersion; Electron beams; Energy measurement; FETs; Scanning electron microscopy; Spectroscopy; Testing; X-ray detection; X-ray detectors;