DocumentCode :
909087
Title :
Calibration Technique for Automated Network Analyzers with Application to Adapter Evaluation
Author :
Engen, Glenn F.
Volume :
22
Issue :
12
fYear :
1974
fDate :
12/1/1974 12:00:00 AM
Firstpage :
1255
Lastpage :
1259
Abstract :
Although conceptually straightforward, the application of existing automated network analyzers to the problem of adapter evaluation is inhibited by the limited accuracy of the detection process, the requirement for several impedance standards at each frequency, and software problems. A hardware modification, which yields improved accuracy for adapter evaluation, is described. An alternative calibration procedure is outlined which exploits this improved accuracy potential, and which requires only one impedance standard.
Keywords :
Application software; Boolean functions; Calibration; Data structures; Detectors; Impedance; NIST; Power measurement; Scattering parameters; Testing;