• DocumentCode
    909087
  • Title

    Calibration Technique for Automated Network Analyzers with Application to Adapter Evaluation

  • Author

    Engen, Glenn F.

  • Volume
    22
  • Issue
    12
  • fYear
    1974
  • fDate
    12/1/1974 12:00:00 AM
  • Firstpage
    1255
  • Lastpage
    1259
  • Abstract
    Although conceptually straightforward, the application of existing automated network analyzers to the problem of adapter evaluation is inhibited by the limited accuracy of the detection process, the requirement for several impedance standards at each frequency, and software problems. A hardware modification, which yields improved accuracy for adapter evaluation, is described. An alternative calibration procedure is outlined which exploits this improved accuracy potential, and which requires only one impedance standard.
  • Keywords
    Application software; Boolean functions; Calibration; Data structures; Detectors; Impedance; NIST; Power measurement; Scattering parameters; Testing;
  • fLanguage
    English
  • Journal_Title
    Microwave Theory and Techniques, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9480
  • Type

    jour

  • DOI
    10.1109/TMTT.1974.1128472
  • Filename
    1128472