DocumentCode
909092
Title
Report on a Position Sensitive Parallel Plate Avalanche Counter with a Distributed Constants Delay-Line-Cathode
Author
Lombardi, M. ; Jilian, Tan ; Potenza, R. ; d´Amico, V.
Author_Institution
I.N.F.N., Laboratori Nazionali di Legnaro Via Romea, 4 - Legnaro (Padova) Italy
Volume
33
Issue
1
fYear
1986
Firstpage
403
Lastpage
406
Abstract
A position sensitive parallel plate (PSPP) stop detector for heavy ions with an active area of 70Ã76 mm2 has been developed. The cathode is a printed circuit board with parallel strips which form a distri buted constant delay-line. The position resolution is better than 1 mm in the x direction (orthogonal to the strips) and 2 cm in the y direction (along the strips). The anode time resolution is about 300 psec fwhm.
Keywords
Anodes; Cathodes; Counting circuits; Delay; Position measurement; Position sensitive particle detectors; Printed circuits; Strips; Timing; Wires;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1986.4337130
Filename
4337130