DocumentCode
909258
Title
Transmission Profile Filtering for Positron Emission Tomography
Author
Palmer, M.R. ; Rogers, J.G. ; Bergstrom, M. ; Beddoes, M.P. ; Pate, B.D.
Author_Institution
UBC/TRIUMF PET Program the University of British Columbia 2211 Wesbrook Mall Vancouver, B.C., V6T 2B5 Canada
Volume
33
Issue
1
fYear
1986
Firstpage
478
Lastpage
481
Abstract
Attenuation correction in PET performed by the transmission measurement method provides accurate compensation but increases image noise. Transmission profile filtering is evaluated in terms of image noise reduction as a trade-off against systematic error introduced by the filter itself. Phantom measurements performed on the UBC/TRIUMF PETT VI tomograph are used to evaluate filter performance. It is found that filtering with a Gaussian of 12.5 mm FWHM reduces the image variance due to noise in transmission profiles by a factor of 15.
Keywords
Aluminum; Attenuation measurement; Councils; Filtering; Filters; Image reconstruction; Imaging phantoms; Noise reduction; Performance evaluation; Positron emission tomography;