• DocumentCode
    909363
  • Title

    Mercuric Iodide X-Ray Camera

  • Author

    Patt, B.E. ; Del Duca, A. ; Dolin, R. ; Ortale, C.

  • Author_Institution
    EG&G Energy Measurements, Inc. Santa Barbara Operations Goleta, California (805) 967-0456
  • Volume
    33
  • Issue
    1
  • fYear
    1986
  • Firstpage
    523
  • Lastpage
    526
  • Abstract
    A prototype x-ray camera utilizing a 1.5-by 1.5-inch, 1024-element, thin mercuric iodide detector array has been tested and evaluated. The microprocessor-based camera is portable and operates at room temperature (Figure 1). Events can be localized within 1-2 mm at energies below 60 keV and within 5-6 mm at energies on the order of 600 keV.
  • Keywords
    Cameras; Electrodes; Energy measurement; Gamma ray detection; Gamma ray detectors; Position sensitive particle detectors; Prototypes; Radiation detectors; Sensor arrays; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1986.4337157
  • Filename
    4337157