DocumentCode
909395
Title
300-Element Silicon-Lithium Position-Sensitive Imaging Detector for Angiography
Author
Walton, J.T. ; Sommer, H.A. ; Thompson, A.C. ; Hughes, E.B. ; Zeman, H.D.
Author_Institution
Lawrence Berkeley Laboratory University of California Berkeley, CA 94720 U.S.A.
Volume
33
Issue
1
fYear
1986
Firstpage
537
Lastpage
541
Abstract
Silicon lithium-drifted [Si(Li)] detectors 150 mm long, 10 mm wide and 5 mm thick with 300 individual elements have been fabricated as imaging detectors for noninvasive studies of human coronary arteries using 33 keV x-rays from a synchrotron radiation beamline. This detector is an extension of earlier work on 30 mm long devices with initially 30 and later 60 elements. The detector fabrication details are discussed highlighting problems in uniform lithium-ion compensation. The device structure is examined and measurements on the interelement impedances presented. Finally an angiograph of the coronary arteries in an excised pig´s heart obtained with this 300-element detector is presented.
Keywords
Angiography; Arteries; Humans; Optical imaging; Position sensitive particle detectors; Radiation detectors; Silicon radiation detectors; Synchrotron radiation; X-ray detection; X-ray detectors;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1986.4337160
Filename
4337160