DocumentCode
911475
Title
Scintillation Photon Detection and Event Selection in High Resolution Positron Emission Tomography
Author
Turko, B.T. ; Zizka, G. ; Lo, C.C. ; Leskovar, B. ; Cahoon, J.L. ; Huesman, R.H. ; Derenzo, S.E. ; Geyer, A.B. ; Budinger, T.F.
Author_Institution
Engineering Division Lawrence Berkeley Laboratory University of California Berkeley, CA 94720 U. S. A. (415) 486-6101
Volume
34
Issue
1
fYear
1987
Firstpage
326
Lastpage
331
Abstract
The scintillation photon detection and event selection subsystem for the high spatial resolution Donner 600-channel positron emission tomograph is described. The tomograph spatial resolution of better than 3.0 mm, FWHM, is obtained by using 600 closely-packed bismuth germanate 3 mm thick scintillation detectors. Each detector has its own 13.5 mm diameter photomultiplier and event selection channel, consisting of two fast pulse amplifiers, a charge integrator, a differential pulse height discriminator and precision time delay for event strobing. The event selection channel is capable of starting a timing cycle from the first photoelectron emitted from the photocathode. It is also capable of discriminating between true events and noise and of eliminating events occurring simultaneously in two adjacent channels. Adjustments of the event strobing time and two discriminator levels are performed by a computer.
Keywords
Bismuth; Event detection; Photomultipliers; Positron emission tomography; Pulse amplifiers; Radioactive decay; Scintillation counters; Single photon emission computed tomography; Solid scintillation detectors; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1987.4337357
Filename
4337357
Link To Document